IEEE-488 Reference
3-152
3.25.8
:DTIMe <NRf>
:TSEQuence:SRESistivity:DTIMe <NRf>
Discharge time; sur resistivity test
:TSEQuence:VRESistivity :DTIMe <NRf>
Discharge time; vol resistivity test
Parameters
<NRf> = 0 to 99999.9 (seconds)
Query
:DTIMe?
Query discharge time
Description
This configuration command is used for the following tests:
Surface Resistivity Test
Volume Resistivity Test
This command is used to specify the discharge time for the specified test.
3.25.9
:PDTime <NRf>
:TSEQuence:SRESistivity:PDTime <NRf>
Pre-discharge time; surface resistivity test
:TSEQuence:VRESistivity:PDTime <NRf>
Pre-discharge time; volume resistivity test
Parameters
<NRf> = 0 to 99999.9 (seconds)
Query
:PDTime?
Query pre-discharge time
Description
This configuration command is used for the following tests:
Surface Resistivity Test
Volume Resistivity Test
This command is used to specify the pre-discharge time for the specified test.
3.25.10 :MVOLtage <NRf>
:TSEQuence:SRESistivity:MVOLtage <NRf>
Measure voltage; sur resistivity test
:TSEQuence:VRESistivity:MVOLtage <NRf>
Measure voltage; vol resistivity test
:TSEQuence:SIResistance:MVOLtage <NRf>
Measure voltage; sur insul resistance test
Parameters
<NRf> = -100.0 to 100.0
Volts; 100V range
-1000 to 1000
Volts; 1000V range
Query
:MVOLtage?
Query measure voltage
Description
This configuration command is used for the following tests:
Surface Resistivity Test
Volume Resistivity Test
Surface Insulation Resistance Test
This command is used to specify the measure voltage for the specified test.