Front Panel Operation
2-33
2.7.1 Resistance measurements
The Model 6517A can make resistance measurements up to
10
17
Ω
using the force voltage measure current (FVMI) tech-
nique. From the known sourced voltage and measured cur-
rent, the Model 6517A calculates and displays the resultant
resistance (R = V/I). The V-Source level can be set automat-
ically by the Model 6517A or it can be manually set by the
user.
The following steps summarize the basic steps to measure re-
sistance:
WARNING
Make sure the V-Source is in standby. In
standby, the OPERATE indicator is off.
The OPER key toggles the V-Source be-
tween standby and operate.
NOTE
To ensure proper operation, always enable
zero check ("ZeroCheck" displayed) be-
fore changing functions (V, I, R, or Q).
The Z-CHK key controls zero check.
1. Enable zero check by pressing Z-CHK.
2. Select RESISTANCE from the MEAS-TYPE selection
of the ohms configuration menu. The ohms configura-
tion menu is accessed by pressing CONFIG and the R
(see paragraph 2.7.3 for details).
NOTE
Step 2 can be skipped if the instrument is
already in the resistance measurement
mode.
3. Select the V-Source adjustment mode. With AUTO V-
Source selected, the instrument will automatically select
the optimum V-Source value (40V or 400V) for the mea-
surement range. With MANUAL V-Source selected, you
select the V-Source range and value. The V-Source ad-
justment mode is selected from the VSOURCE item of
the CONFIGURE OHMS menu. See paragraphs 2.7
(Auto V-Source) and 2.7.3 (VSOURCE) for details.
4. Connect the resistance to be measured to the Model
6517A. Figure 2-31 shows typical connections while
Figure 2-32 shows connections using the Model 8002A
High Resistance Test Fixture.
NOTE
The connections in Figure 2-33 assume
that V-Source LO is internally connected
to ammeter LO. This internal connection
is controlled from the METER-CON-
NECT option of the CONFIGURE V-
SOURCE menu (see paragraph 2.9.1).
This LO-to-LO connection can instead be
made by using an external cable to connect
V-Source LO to ammeter LO.
5. Select the ohms function by pressing the R key.
6. If the manual V-Source adjustment mode is selected, use
the
,
, and the VOLTAGE SOURCE
▲
and
▼
keys to set the voltage level. The V-Source range can be
changed from the RANGE item of the CONFIGURE V-
SOURCE menu. See paragraph 2.9.2 for details on set-
ting range and level for the V-Source. Note that you will
not be able to adjust the V-Source if AUTO V-Source is
selected.
WARNING
To avoid a possible shock hazard, do not
use a voltage level that exceeds the max-
imum input voltage rating of the test fix-
ture. For example, the maximum input
voltage to the Model 8002A High Resis-
tance Test Fixture must not exceed
200V peak.
7. Use the
▲
and
▼
RANGE keys to select the resistance
measurement range, or select AUTO range. Note that
with AUTO range selected, the instrument will not go to
the 2T
Ω
, 20T
Ω
and 200T
Ω
ranges.
NOTE
For optimum accuracy, leakage currents in
the test fixture can be cancelled by per-
forming REL on the current component of
the measurement. To cancel leakage cur-
rent, perform “Cancelling Test Fixture
Leakage Current” which follows Step 9 of
this procedure.
Table 2-11
Ohms reading ranges and AUTO V-Source
Reading range
AUTO V-Source
Test
voltage
Amps
range
200k
Ω
– 2M
Ω
2M
Ω
– 20M
Ω
20M
Ω
– 200M
Ω
200M
Ω
– 2G
Ω
2G
Ω
– 20G
Ω
20G
Ω
– 200G
Ω
200G
Ω
– 2T
Ω
2T
Ω
– 20T
Ω
20T
Ω
– 200T
Ω
40V
40V
40V
40V
40V
40V
400V
400V
400V
200µA
20µA
2µA
200nA
20nA
2nA
2nA
200pA
20pA