![background image](http://html2.mh-extra.com/html/keithley/6517a/6517a_user-manual_4066185174.webp)
IEEE-488 Reference
3-23
:SYSTem:PRESet
Long-form
:SYST:PRES
Short-form
:SYSTem:PRES
Long and short-form combination
Note that each command word must be in long-form or short-form, and not something in
between. For example, :SYSTe:PRESe is illegal and will generate an error. The command
will not be executed.
Short-form rules:
The following rules can be used to determine the short-form version of
any SCPI command or parameter:
A. If the length is four letters or less, there is no short-form version. Example:
:auto = :auto
B. The following rules apply to words that exceed four letters:
a. If the fourth letter of the word is a vowel, delete it and all the letters after it. Exam-
ple:
:immediate = :imm
Note: The “y” in POLynomial is a vowel. Thus, polynomial = pol.
b. If the fourth letter of the word is a consonant, retain it but drop all the letters after
it. Example:
:output = :outp
C. If there is a question mark (?; query) or a non-optional number included in the com-
mand word, it must be included in the short- form version. Examples:
:delay? = :del? :layer2 = :lay2
D. As previously explained, command words or characters that are enclosed in brackets
([]) are optional and need not be included in the program message.