Front Panel Operation
2-39
2.7.3 Ohms configuration
The following information explains the various configura-
tion options for the ohms function. The configuration menu
is summarized in Table 2-12. This menu is accessed by press-
ing CONFIG and then R. Paragraph 2.3.5 summarizes the
rules for navigating through the menu structure.
Note that a function does not have to be selected in order to
be configured. When the function is selected, it will assume
the programmed status.
SPEED
The SPEED parameter sets the integration time of the A/D
converter, the period of time the input signal is measured (al-
so known as aperture). It is discussed in paragraph 2.5.2.
FILTER
Use this menu item to configure the two basic filter types; av-
eraging and median. Note that you can use either the averag-
ing filter, the median filter, or both.
The filter menu is available from the function configuration
menus (i.e. press CONFIG V) or by pressing CONFIG FIL-
TER with the desired function already selected. All of the pa-
Model 6517A
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
METER
SOURCE
LID
INTERLOCK
TRIAX
250 MAX
HI-LO
MAX INPUT
1100V
!
INPUT
250V PEAK
COMMON
OUT
!
LINE RATING
90-134VAC
180-250VAC
50, 60, 400HZ
55VA MAX
LINE F
SLOWB
1/2A, 2
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
PREAMP OUT
250V PEAK
V SOURCE
LO
HI
INTERLOCK
!
Warning: Connect of fixture
to safety earth ground using
safety ground wire (supplied
with 8002A test fixture).
7078-TRX-3 Triax Cable
8607 Banana Plug Cables
6517-ILC-3 Interlock Cable
Model 8009
Figure 2-36
Connections for measurements using Model 8009 test fixture
rameters (menu items) for FILTER are explained in
paragraph 2.17.
RESOLUTION
The RESOLUTION parameter sets the display resolution. It
is discussed in paragraphs 2.5.2 and 2.12.
AMPSREL
Leakage current in a test fixture can corrupt a resistance mea-
surement. This leakage current can be cancelled by perform-
ing a REL on the current component of the measurement.
With this menu item, you can use the established amps REL
value for the resistance measurement. See “Cancelling Test
Fixture Leakage Current” in paragraph 2.7.1.
ENABLED: Use this option to use the amps REL value. Af-
ter this option is selected, the instrument will display the sta-
tus of REL for the ohms function and for the amps function.
If REL for the amps function is disabled, then no amps REL
operation will be performed on the measurement.
DISABLED: Use this option if you do not wish to use the
amps REL value for resistance measurements.