Section 13: LPT library function reference
Model 4200A-SCS Parameter Analyzer Reference Manual
13-14
4200A-901-01 Rev. C / February 2017
To abort a running
pulse_exec
pulse test, see
dev_abort
.
devint
does the following:
1.
Clears all sources by calling
devclr
.
2.
Clears the matrix crosspoints by calling
clrcon
.
3.
Clears the trigger tables by calling
clrtrg
.
4.
Clears the sweep tables by calling
clrscn
.
5.
Resets GPIB instruments by sending the string defined with
kibdefint
.
6.
Resets the active instrument cards.
Instrument cards are reset in the following order:
1.
SMU instrument cards
2.
CVU instrument cards
3.
Pulse instrument cards (4225-PMU or 4220-PGU)
devint
is implicitly called by
execut
and
tstdsl
.
devclr
is implicitly called by
clrcon
.
The SMUs return to the following states:
•
100 µA and 10 V ranges
•
Autorange on
•
Voltage source
•
0 V DCV bias
The 4210-CVU returns to the following
states:
•
30 mV
RMS
AC signal
•
0 V DCV bias
•
100 kHz frequency
•
Autorange on
•
Cable length compensation set to 0 m
•
Open/Short/Load compensation disabled