Model 4200A-SCS Parameter Analyzer Reference Manual
Section 4: Multi-frequency capacitance-voltage unit
4200A-901-01 Rev. C / February 2017
4-99
c-2vsv-solarcell test
Make sure the solar cell is connected to the 4210-CVU (see
(on page 4-5)).
This test uses a CVU to perform a capacitance-voltage sweep on a solar cell. Capacitance is
measured on each step of the sweep. A graph (1/C
2
and C versus V) is generated from the collected
data. The Formulator calculates 1/C
2
and the doping density.
c-2vsv-solarcell Analyze sheet
The test data is displayed in the Analyze sheet:
•
Cp_AC: Measured parallel capacitance.
•
Gp_AC: Measured conductance.
•
DCV_AC: Forced DC bias voltage.
•
F_AC: Forced test frequency.
•
CVU1S: Status code for each measurement. Rows highlighted in blue indicate a fault. For details,
see
(on page 6-191).
•
Formulas: Formulator calculation results.
AC = Anode to cathode.
cfsweep test
Make sure the PV cell is connected to the 4210-CVU (see
(on page 4-5)).
This test performs a frequency sweep, measuring capacitance at each frequency point, and then
generates a C versus F graph.
cfsweep Analyze sheet
The test data is displayed in the Analyze sheet:
•
Cp_AC: Measured parallel capacitance.
•
Gp_AC: Measured conductance.
•
DCV_AC: Forced DC bias voltage.
•
F_AC: Forced test frequency.
•
CVU1S: Status code for each measurement. Rows highlighted in blue indicate a fault. For details,
see
(on page 6-191).