Enables or disables the analog filter:
•
Disable:
0
•
Enable (can minimize the amount noise in the readings, but increases
measurement time):
1
ReadingRate
The reading rate used to acquire the measurements (0 to 4); see
Details
C
Output: The measured capacitance
V
Output: The bias voltage used
G_or_R
Output:
•
Parallel measurement model:
G_or_R
is the measured conductance
•
Series measurement model:
G_or_R
is the measured resistance
Details
This user module is used to make a single, fixed-bias capacitance and conductance measurement.
The default parameters for the
590-cmeas
UTM, which uses the
Cmeas590
user module, are shown
in the figure below.
Figure 560: Cmeas590 (cmeas UTM parameters)
In general, the 590 is set to source 5 V for 2 seconds, then make a measurement, as shown in the
following figure.
Figure 561: Cmeas590 measurement