Model 4200A-SCS Parameter Analyzer Reference Manual
Section 5: Pulse measure and pulse generator units
4200A-901-01 Rev. C / February 2017
5-33
Enabling connection compensation
This procedure is for ITMs. For UTMs, you need to enable connection compensation data using the
(on page 13-96) function.
To apply the connection compensation data to DUT measurements, you must enable connection
compensation for the test.
To enable connection compensation:
1. Select the test.
2. Select
Configure
.
3. Select the terminal to be compensated.
4. In the right pane, select
Terminal Settings
.
5. Select
Advanced
. The PMU Advanced Settings dialog box is displayed, as shown in the next
figure.
Figure 172: Enabling connection compensation
6. To enable connection compensation, select
Short Connection
and click
OK
.
7. To disable disable connection compensation, clear
Short Connection
. When disabled,
connection compensation values are not applied to DUT measurements.