Model 4200A-SCS Parameter Analyzer Reference Manual
Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017
9-31
LI
This command enables voltage and current functions to be measured when the 4200A-SCS is in list display
mode.
Usage
LI '
AAAAAA
'
LI '
AAAAAA
', '
AAAAAA
'
LI '
AAAAAA
', '
AAAAAA
', '
AAAAAA
'
LI '
AAAAAA
', '
AAAAAA
', '
AAAAAA
', '
AAAAAA
'
LI '
AAAAAA
', '
AAAAAA
', '
AAAAAA
', '
AAAAAA
', '
AAAAAA
'
LI '
AAAAAA
', '
AAAAAA
', '
AAAAAA
', '
AAAAAA
', '
AAAAAA
', '
AAAAAA
'
AAAAAA
A name assigned for
CH
,
VS
, or
VM
; see channel definition page (
DE
)
Details
When the 4200A-SCS is in the list display mode (
DM2
asserted), the
LI
command enables functions
to be measured in a test sequence. To enable a function, include the SMU channel name (as
assigned by the
CH
command), voltage source name (as assigned by the
VS
command), or the
voltmeter name (as assigned by the
VM
command) in the command string. The
DE
page is used to
assign names to voltage and current functions.
Only functions that are specified (enabled) are measured. Data sheet columns for disabled functions
are not shown.
Example
L1 'V1', 'I1', 'VS1'
Assume the following names have been assigned using the
DE
page for SMU channel 1 (
CH1
):
•
Voltage is named
V1
•
Current is named
I1
•
Voltage source is named
VS1
The command string enables the above functions for a test sequence.
Also see
(on page 9-30)