Section 8: Keithley User Library Tool (KULT)
Model 4200A-SCS Parameter Analyzer Reference Manual
8-4
4200A-901-01 Rev. C / February 2017
Parameters tab area
In the Parameters tab, you define and display parameters in the user module call. You can define and
display:
•
Parameter name
•
Parameter data type
•
Input or output (I/O) data direction
•
Default, min, and max values for the parameter
These options are defined in the following text.
The Parameters tab area is near the bottom of the KULT main screen. An example is shown here.
Figure 420: Parameters tab for the Rdson42XX user module from the KI42XX library
You can right-click anywhere in the Parameters tab area to access the Add, Delete, and Apply
options.
To add a parameter:
1. Click
Add
.
2. Enter the information as needed.
3. Click
Apply
.
To delete a parameter:
1. Click the parameter name or any of the adjacent fields.
2. Click
Delete
.
To make changes to the parameters:
1. Make changes in the appropriate field.
2. Click
Apply
.
Parameter name field
The parameter name field identifies the parameters that are passed to the user module. These are
the same parameters that are specified in the user-module function prototype (which KULT constructs
from the Parameters tab entries when you click
Apply,
and then displays in the module-parameter
display area).