VC and IC
This command configures the SMU to output a fixed (constant) voltage or current level.
Usage
AAB
, ±
CCC.CCCC
, ±
DDD.DDDD
AA
Source mode:
•
Voltage:
VC
•
Current:
IC
B
SMU channel number: 1 to 9
CCC.CCCC
Output value:
•
Voltage source:
−
210.00 V to +210.00 V
•
Current source, 4200-SMU:
−
0.1050 A to +0.1050 A
•
Current source, 4210-SMU:
−
1.0500 A to +1.0500 A
DDD.DDDD
Compliance value:
•
Voltage source:
−
210.00 V to +210.00 V
•
Current source, 4200-SMU:
−
0.1050 A to +0.1050 A
•
Current source, 4210-SMU:
−
1.0500 A to +1.0500 A
Details
For any channel configured as a SMU, this command configures the SMU to output a fixed (constant)
voltage or current level.
With the voltage source mode (
VC
) specified in the command string, you set the current compliance.
For the current source mode (
IC
), you set the voltage compliance.
Also see
Keithley Configuration Utility (KCon)