Section 12: Maintenance
Model 4200A-SCS Parameter Analyzer Reference Manual
12-6
4200A-901-01 Rev. C / February 2017
5. Select
SMU Auto Calibration
. A warning dialog box is displayed.
6. Select
OK.
The SMU Auto Calibration dialog box opens, as shown in the following figure.
Figure 491: SMU Auto Calibration dialog box
7. Select
Start
. A progress bar is displayed.
When auto calibration is complete, the message "Auto calibration successfully completed" is
displayed.
8. Select
Close
. Auto calibration is complete.