Model 4200A-SCS Parameter Analyzer Reference Manual
Section 5: Pulse measure and pulse generator units
4200A-901-01 Rev. C / February 2017
5-21
The preview displays representations of the waveforms for the different PMU channels in the test.
The illustrated pulse waveforms are defined by the Pulse Timing settings for the specific PMU. This
graphical depiction provides a representation of parameter values (no signals are output to the test
device). It is not intended to be a strict definition of the actual number of pulses or pulse voltages that
will be applied to the device under test (DUT). When the actual test is running, the actual number of
pulses or pulse voltages may be different from the preview.
Enabling current measure ranging or load-line effect compensation (LLEC) requires additional pulses
to what is shown in the preview. To learn about how the PMU handles device under test (DUT) load
variation and measure ranging during a test, see
How LLEC adjusts pulse output to the target levels
(on page 5-35).
To open the PMU Advanced Test Settings dialog box, select
Test Settings
, then
Advanced
(see
next figure).
Figure 160: Timing button (PDU Definition tab)
Review the following examples to understand the Pulse Timing Preview feature.