Section 9: Keithley External Control Interface (KXCI)
Model 4200A-SCS Parameter Analyzer Reference Manual
9-76
4200A-901-01 Rev. C / February 2017
:CVU:LENGTH
This command selects the cable length for the specified CVU card.
Usage
:CVU:LENGTH
len
len
The cable length:
•
0
: 0 m (no cable compensation)
•
1.5
: 1.5 m CVU cable
•
3.0
: 3.0 m CVU cable
•
4.0
: Custom
•
5.0
: 1.5 m CVIV cable; 2-wire mode
•
6.0
: 1.5 m CVU to CVIV cable with 0.75 m CVIV to DUT cable; 4-wire mode
•
7.0
: Blue 1.5 m CVU to CVIV cable with 0.61 m CVIV to DUT cable; 4-wire
mode
Also see
None
:CVU:CORRECT
This command enables or disables open, short, and load correction for the specified CVU card.
Usage
:CVU:CORRECT
open
,
short
,
load
open
off:
0
on:
1
short
off:
0
on:
1
load
off:
0
on:
1
Details
Each of the parameters is a Boolean (
0
= off,
1
= on).
Also see
None
:CVU:CABLE:COMP:OPEN
This command performs open compensation and collects the open compensation cable data for the CVU.
Usage
:CVU:CABLE:COMP:OPEN
Also see
(on page 4-10)