Model 4200A-SCS Parameter Analyzer Reference Manual
Section 8: Keithley User Library Tool (KULT)
4200A-901-01 Rev. C / February 2017
8-31
10. In the 5th line, enter the following forced-voltage parameter information:
Parameter Name:
Vforce
Data type:
D_ARRAY_T
I/O: Output
11. In the 6th line, under
Parameter Name
, change
ArrSizeForParm5
to
NumVPoints
.
12. In the 6th line, under
Default
, enter the number
11
for the default voltage array size.
When executing the
VSweep
user module in a UTM, the current and voltage array sizes must match;
NumIPoints
must equal
NumVPoints
. If the sizes do not match, the second return statement in the
code halts execution and returns an error number (
-2
) in the
VSweep
column of the UTM Data
worksheet.
13. Click
Apply.
In the module-parameter display area, the function prototype now includes the
declared parameters, as shown below.
Figure 441: VSweep user-module dialog box after entering and applying code and parameters
Entering the VSweep user-module header files
(on page 8-32).