15-10
SCPI Reference Tables
Model 2750 Multimeter/Switch System User’s Manual
[SENSe[1]]
:VOLTage:AC
Path to configure AC voltage.
✓
:APERture <NRf>
[, <clist>]
Set integration rate in seconds (60Hz,
1.67e-4 to 1; 50Hz, 2e-4 to 1).
(Note 2)
:APERture? [<clist>]
Query aperture integration rate.
:NPLCycles <n>[, <clist>]
Set integration rate in line cycles (60Hz,
0.01 to 60; 50Hz, 0.01 to 50).
5.0
✓
:NPLCycles? [<clist>]
Query line cycle integration rate.
✓
:RANGe
Path to set measurement range:
✓
[:UPPer] <n> [, <clist>]
Select range (0 to 757.5).
750
✓
[:UPPer]? [<clist>]
Query range.
✓
:AUTO <b> [, <clist>]
Enable or disable auto range.
ON
✓
:AUTO? [<clist>]
Query state of auto range.
✓
:DIGits <n> [, <clist>]
Specify measurement resolution (4 to 7).
6
:DIGits? [<clist>]
Query resolution.
:REFerence <n> [, <clist>]
Specify reference (-757.5 to 757.5).
0
✓
:STATe <b> [, <clist>]
Enable or disable reference.
OFF
✓
:STATe? [<clist>]
Query state of reference.
✓
:ACQuire [, <clist>]
Use input signal as reference.
:REFerence? [<clist>]
Query reference value.
✓
:AVERage
Path to configure and control filter:
:TCONtrol <name>
Select filter type: (MOVing or REPeat).
(Note 3)
:TCONtrol?
Query filter type.
:WINDow <NRf>
Set filter window in % of range
(0 to 10).
0.1
:WINDow?
Query filter window.
:COUNt <n> [, <clist>]
Specify filter count (1 to 100).
10
:COUNt? [<clist>]
Query filter count.
[:STATe] <b> [, <clist>]
Enable or disable filter.
(Note 4)
[:STATe]? [<clist>]
Query state of digital filter.
:DETector
Path to set bandwidth:
:BANDwidth <NRf>
[, <clist>]
Set AC detector bandwidth in Hertz
(3 to 3e5).
30
:BANDwidth? [<clist>]
Query bandwidth.
Table 15-5 (continued)
SENSe command summary
Command
Description
Default
parameter
Ref
SCPI
2750-900-01.book Page 10 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...