REN (remote enable)
The remote enable command is sent to the Model 2750 by the controller to set up the
instrument for remote operation. Generally, the instrument should be placed in the remote
mode before you attempt to program it over the bus. Setting REN true does not place the
instrument in the remote state. You must address the instrument to listen after setting REN
true before it goes into remote.
The Model 2750 must be in remote in order to use the following commands to trigger and
acquire readings:
•
INITiate and then FETCh?
•
READ?
•
MEASure?
IFC (interface clear)
The IFC command is sent by the controller to place the Model 2750 in the local, talker, lis-
tener idle states. The unit responds to the IFC command by canceling front panel TALK or
LSTN lights, if the instrument was previously placed in one of these states.
Note that this command does not affect the status of the instrument. Settings, data, and
event registers are not changed.
With auto output off enabled (SOURce1:CLEar:AUTO ON), the output will remain on if
operation is terminated before the output has a chance to automatically turn off.
To send the IFC command, the controller need only set the IFC line true for a minimum of
100µs.
LLO (local lockout)
Use the LLO command to prevent local operation of the instrument. After the unit receives
LLO, all of its front panel controls except OUTPUT OFF are inoperative. In this state,
pressing LOCAL will not restore control to the front panel. The GTL command restores
control to the front panel. Cycling power will also cancel local lockout.
GTL (go to local)
Use the GTL command to put a remote-mode instrument into local mode. The GTL com-
mand also restores front panel key operation.
2750-900-01.book Page 8 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...