Triggering
8-11
C.
For the first pass through the model, the scanner does not wait at point B for a trig-
ger. Instead, it closes the first channel.
D.
After the relay settles, the Model 7002 outputs a Channel Ready pulse. Since the
instrument is programmed to scan 400 channels, operation loops back up to point
B, where it waits for an input trigger.
E & F.
Model 2750 operation is at point A waiting for a trigger. The output Channel
Ready pulse from the Model 7002 triggers the Model 2750 to measure DUT #1
(point E). After the measurement is complete, the Model 2750 outputs a comple-
tion pulse (point F) and then loops back to point A where it waits for another input
trigger.
The trigger applied to the Model 7002 from the Model 2750 closes the next channel in the
scan. This triggers the Model 2750 to measure the next DUT. The process continues until
all 400 channels are scanned, measured, and stored in the buffer.
External triggering with BNC connections
An adapter cable is available to connect the micro-DIN Trigger Link of the Model 2750 to
instruments with BNC trigger connections. The Model 8503 DIN to BNC Trigger Cable
has a micro-DIN connector at one end and two BNC connectors at the other end. The BNC
cables are labeled VMC (trigger line 1) and EXT TRIG (trigger line 2).
shows how a Keithley Model 220 Current Source can be connected to the Trig-
ger Link of the Model 2750 using the adapter cable. When used with the STEP mode of
the Model 220, you can perform synchronized source-measure operations without the use
of a computer. Whenever the Model 220 receives a trigger from the Model 2750, it will
step to the next current source value.
2750-900-01.book Page 11 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...