Model 2750 Multimeter/Switch System User’s Manual
Range, Digits, Rate, Bandwidth, and Filter
4-13
To set bandwidth, simply specify (approximately) the frequency of the input signal. The
instrument will automatically set the bandwidth as follows:
<NRf>
= 3 to 29
3Hz to 300kHz
= 30 to 299
30Hz to 300kHz
= 300 to 300e3
300Hz to 300kHz
NOTE
A rate command (:NPLCycles or :APERture) for ACV and ACI is only valid if
the bandwidth for that AC function is set to 300 (300Hz to 300kHz). See “
” for details.
Rate and bandwidth conflict error
For bandwidth settings of 3 and 30, the normal A/D conversion method is not used for
ACV and ACI measurements. Therefore, integration rate commands (:NPLCycles and
:APERture) for these bandwidth settings will cause a settings conflict error (-221) and not
be executed.
For a bandwidth setting of 300, the normal A/D conversion method is used, therefore the
integration rate commands can be used.
The default bandwidth setting is 30 for both ACV and ACI. If you want to set an integra-
tion rate for an AC function, you will have to first set the bandwidth to 300.
Rate and bandwidth programming examples
Example #1
— The following command sequence sets ACV rate to 5 PLC. In order to set
rate for an AC function, bandwidth must first be set to 300:
VOLT:AC:DET:BAND 300
' Set ACV bandwidth to 300.
VOLT:AC:NPLC 5
' Set ACV rate to 5 PLC.
NOTE
VOLT:AC:DET:BAND must be set to 300 before the VOLT:AC:NPLC command
can be sent.
Example #2
— The following command sequence configures channels 101 and 103 of the
Model 7700 to set integration rate to 6 PLC when they are scanned.
FUNC 'VOLT' (@101, 103)
' Select DCV function.
VOLT:NPLC 6,(@101, 103)
' Set rate to 6 PLC.
2750-900-01.book Page 13 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...