![background image](http://html1.mh-extra.com/html/keithley/2750/2750_user-manual_3956564115.webp)
3-24
Basic DMM Operation
Model 2750 Multimeter/Switch System User’s Manual
NOTE
The states (on or off) of dry circuit ohms and offset-compensated ohms are
“remembered” by the
Ω
4 function after you select a different measurement func-
tion (i.e., DCV). When
Ω
4 is again selected, the previous states of dry circuit
ohms and offset-compensated ohms will be restored.
The accuracy specifications for dry circuit ohms is for offset-compensated ohms
and line synchronization enabled.
Dry circuit ohms measurement considerations
Dry circuit ohms uses a constant current source with voltage monitoring that is used to
clamp the current source voltage. The current source will remain constant as long as the
monitoring voltage is <20mV. When the voltage exceeds 20mV, the current source shunts
current internal to the DMM until 20mV is maintained at the DUT.
The voltage is monitored at the Sense HI terminal. This allows the Model 2750 to accom-
modate an additional 80% of range of path resistance per lead. So when measuring a 1
Ω
resistor through a switching module, an additional 0.8
Ω
per lead of path resistance can be
tolerated. Lead resistance of several 100m
Ω
is common with a switching module, espe-
cially when using ribbon cable wire.
There is a secondary voltage monitoring circuit on Input HI. This is needed if Sense HI or
Input HI have different voltages. This could occur if measuring 4-wire resistance and
Input HI becomes an open circuit. In this condition, Sense HI would indicate no voltage
and the secondary voltage monitor would clamp the current source to <50mV.
When using dry circuit ohms, the DUT is shunted by 100k
Ω
and 0.9µF. This allows the
current source to have minimal overshoot voltage under transient conditions. When used
with a switching system, the overshoot is <70mV in 25µsec.
2750-900-01.book Page 24 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...