Model 2750 Multimeter/Switch System User’s Manual
Getting Started
1-29
Figure 1-6
Connection to DMM for 4-wire function (system channel 106 closed)
NOTE
Switching module channels can also be controlled using multiple channel opera-
tion. This allows individual control of all module channels (switches). Multiple
channel operation should only be used by experienced service personnel who
recognize the dangers associated with multiple channel closures. See
for details.
Close/open operation
The following points on operation pertain to system channel operation only:
•
Only one input channel (or channel pair) is closed at one time. When you close an
input channel, the previously closed input channel(s) will open.
•
When a system channel is closed, the channel number will be displayed on the
Model 2750. The slot number for the module is also displayed. For example, “103”
indicates that system input channel 3 for a module in slot 1 is closed.
•
The paired channel for a 4-wire function is not displayed. Only the system channel
number is displayed. For example, in
, channel number 106 will be dis-
played with the Model 7707 installed in slot 1 of the mainframe.
•
Switching modules that have current measurement capability have separate chan-
nels reserved exclusively for the DCI and ACI functions. For example, the Model
7700 has channels 21 and 22 reserved for amps measurements. With the DCI or
ACI function selected, only channels 21 and 22 can be closed. These channels can-
not be accessed on any other function.
shows the front panel keys used to close and open system channels.
DUT
DMM
DMM
7700
Switching Module
7700
Switching Module
HI
LO
HI
LO
Input
Ch 6
HI
LO
HI
LO
Sense
Ch 16
HI
LO
Input
HI
LO
HI
LO
Sense
Ch 16
HI
LO
Ch 6
2750-900-01.book Page 29 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...