Model 2750 Multimeter/Switch System User’s Manual
Rel, Math, Ratio, Channel Average, dB
5-13
Remote programming — math
Math commands
NOTE
When measurements are performed, the readings are fed to other enabled pro-
cessing operations, including Math.
explains “
” and the commands used to return Math results.
The commands to perform math calculations are listed in
. Details on these com-
mands follow the table.
NOTE
Queries are not included in
. All the math commands are provided in
.
Table 5-2
Math commands
Commands
1
Description
Def
CALCulate[1]:FORMat <name> [, <clist>]
Select calculation; <name> = NONE,
MXB, PERCent, or RECiprocal.
PERC
CALCulate[1]:KMATh:MMFactor <NRf>
[, <clist>]
Set mX+b “m” factor; <NRf> =
-4294967295 to +4294967295.
1
CALCulate[1]:KMATh:MBFactor <NRf>
[, <clist>]
Set mX+b “b” factor; <NRf> =
-4294967295 to +4294967295.
0
CALCulate[1]:KMATh:MUNits <char> [, <clist>]
Set mX+b units; see “Setting mX+b
units.”
‘X’
CALCulate[1]:KMATh:PERCent <NRf> [, <clist>] Set reference value for percent; <NRf> =
-4294967295 to +4294967295.
1
CALCulate[1]:KMATh:PERCent:ACQuire
Use input signal as reference value.
CALCulate[1]:STATe <b> [, <clist>]
Enable or disable calculation; <b> = ON
or OFF.
(Note 2)
CALCulate[1]:DATA[:LATest]?
Return last result of calculation.
CALCulate[1]:DATA:FRESh?
Return last “fresh” result of calculation.
Channel list parameter:
<clist> = (@SCH)
where: S
= Mainframe slot number (1, 2, 3, 4 or 5)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
1. The <clist> parameter is used to configure one or more channels for a scan.
2. *RST default is ON, SYSTem:PRESet is OFF.
2750-900-01.book Page 13 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...