Model 2750 Multimeter/Switch System User’s Manual
Basic DMM Operation
3-25
Temperature measurements
The Model 2750 can measure temperature using thermocouples, thermistors, and 4-wire
RTDs. When deciding which temperature sensor to use, keep in mind that the thermocou-
ple is the most versatile, the thermistor is the most sensitive, and the 4-wire RTD is the
most stable.
Thermocouples
For thermocouples, temperature measurement range depends on which type of thermocou-
ple is being used. Thermocouples that are supported include types J, K, N, T, E, R, S, and
B.
When two wires made up of dissimilar metals are joined together, a voltage is generated.
The generated voltage is a function of temperature. As temperature changes, the voltage
changes. The thermocouple voltage equates to a temperature reading. This is the basic
operation principle of the thermocouple.
NOTE
The equation to calculate thermocouple temperature is provided in
When you connect a thermocouple directly to the input of the Model 2750, at least one of
those connections will be a junction made up of two dissimilar metals. Hence, another
voltage is introduced and is algebraically added to the thermocouple voltage. The result
will be an erroneous temperature measurement.
To cancel the affects of the unwanted thermal voltage, the thermocouple circuit requires a
reference junction that is at a known temperature.
Type
Range
Resolution
J
-200°C to 760°C
0.001°C
K
-200°C to 1372°C 0.001°C
N
-200°C to 1300°C 0.001°C
T
-200°C to 400°C
0.001°C
E
-200°C to 1000°C 0.001°C
R
0°C to 1768°C
0.1°C
S
0°C to 1786°C
0.1°C
B
+350°C to 1820°C 0.1°C
2750-900-01.book Page 25 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...