13-8
SCPI Signal Oriented Commands
Model 2750 Multimeter/Switch System User’s Manual
When a MEASure? command is sent, the specified function is selected.
If specified, range and resolution will also set. All other instrument set-
tings related to the selected function are reset to the *RST defaults.
NOTE
If a function is not specified, the command executes as if the present function is
specified. For example, assume the
Ω
2 function is presently selected. When
MEAS? is sent, the instrument resets to the *RST defaults for the
Ω
2 function,
and then performs a measurement.
When this command is sent, the following commands execute in the
order that they are presented:
ABORt
CONFigure:<function>
READ?
When ABORt is executed, the instrument goes into the idle state if con-
tinuous initiation is disabled. If continuous initiation is enabled, the
operation re-starts at the beginning of the Trigger Model.
When CONfigure is executed, the MEASure? parameters (<rang>,
<res> and <clist>) are executed and the instrument goes into the “one-
shot” measurement mode. It is similar to sending the CONFigure com-
mand with no <clist> parameter. See
“CONFigure:<function> [<rang>],
[<res>], [<clist>],” page 13-3
, for more details.
When READ? is executed, its operations will then be performed. In
general, another ABORt is performed, then an INITiate, and finally a
FETCh? to acquire the reading. See
, for more
details.
Programming examples:
Programming example #1
— The following command measures DCV
on channel 101 using the 10V range with 3
H
digit display resolution:
MEAS:VOLT? 10, 0.01, (@101)
Programming example #2
— The following command measures DCV
on the 100V range:
MEAS:VOLT? 100
2750-900-01.book Page 8 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...