![background image](http://html1.mh-extra.com/html/keithley/2750/2750_user-manual_3956564225.webp)
7-20
Scanning
Model 2750 Multimeter/Switch System User’s Manual
NOTES
Auto channel configuration cannot be enabled if there is a non-scan channel
presently closed. For example, assume the scan list consists of channels 105
through 110, and channel 101 (a non-scan channel) is presently closed. When
you attempt to enable auto channel configuration from the front panel, the mes-
sage “NOT IN SCAN” is briefly displayed. For remote operation, error -221
(settings conflict) occurs.
With auto channel configuration enabled, the
and
keys will not properly
step through a non-sequential scan list. Therefore, auto channel configuration
should not be used for a non-sequential scan list. For information on non-
sequential scans, see “
.
For remote operation, the ROUT:CLOS:ACON command is used to enable or
disable auto channel configuration (
).
Saving setup
Up to three instrument setups can be saved in memory using the SHIFT > SAVE menu
(SAV0, SAV1, and SAV2). A user-saved setup can also be used as the power-on setup. A
user-saved setup can be restored from the SHIFT > SETUP menu. Details on user-setups
are covered in
.
Auto scan
When auto scan is enabled, the scan operation is saved in memory. If power to the
Model 2750 is interrupted, the scan will resume when power is restored. With auto scan
enabled, the last scan setup becomes the power-on setup. It takes precedence over the fac-
tory, *RST or user-saved power-on setup.
Perform the following steps to set auto scan.
1.
While in the normal measurement state, press SHIFT and then SETUP.
2. Use the
Δ
and
∇
keys to display the auto scan (AUTOSCAN) setting; N (no) or Y
(yes).
3.
To change the setting, press
to place the cursor on the setting (N or Y) and press
Δ
or
∇
to change the setting.
4.
Press ENTER to exit from the menu structure.
NOTE
With auto scan enabled, DO NOT save the present setup as the power-on default
setup. If you do so, an interrupted scan will not resume.
If during the power-up sequence the Model 2750 detects a card ID change for
any slot, auto scan configuration will disable and an interrupted scan will not be
2750-900-01.book Page 20 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...