14-2
FORMat and Misc SYSTem Commands
Model 2750 Multimeter/Switch System User’s Manual
FORMat commands
The commands in this subsystem are used to select the format for transferring data,
, over the bus.
FORMat[:DATA] <type>[,<length>]
Parameters
ASCii
= ASCII format
SREal
= Binary IEEE-754 single precision format
REAL, 32
= Binary IEEE-754 single precision format
DREal
= Binary IEEE-754 double precision format
REAL, 64
= Binary IEEE-754 double precision format
NOTE
<length> is not used for the ASCii, SREal, or DREal parameters.
The response to READ?, FETCh?, MEASure?, TRACe:DATA?, CALC1:DATA?, or
CALC2:DATA? over the GPIB can be returned in either the ASCII or binary format. All
other queries are returned in ASCII, regardless of the selected format. Over the RS-232
interface, only the ASCII format is allowed.
NOTE
Regardless of which data format for output strings is selected, the instrument
will only respond to input commands using the ASCII format.
Table 14-1
SCPI commands — data format
Command
Description
Default
FORMat[:DATA] <type>[,<length>]
Specify data format; ASCii, SREal, REAL, 32,
DREal, or REAL, 64.
ASCii
FORMat:ELEMents <item list>
Specify data elements; READing, UNITs, TSTamp,
RNUMber, CHANnel, LIMits.
All 3
FORMat:BORDer <name>
Specify byte order; NORMal or SWAPped.
(see Note)
Note: *RST default is NORMal. SYSTem:PRESet default is SWAPped.
2750-900-01.book Page 2 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...