Model 2750 Multimeter/Switch System User’s Manual
Rel, Math, Ratio, Channel Average, dB
5-5
Commands
1
Description
Default
Rel commands for ACI
[SENSe[1]]
Optional root command.
:CURRent:AC:REFerence <n> [, <clist>]
Specify rel value; <n> = -3.1 to 3.1 (A).
0
:CURRent:AC:REFerence:STATe <b>
[, <clist>]
Enable/disable rel; <b> = ON or OFF.
OFF
:CURRent:AC:REFerence:ACQuire [, <clist>]
Use input signal as rel value.
Rel commands for
Ω
2
[SENSe[1]]
Optional root command.
:RESistance:REFerence <n> [, <clist>]
Specify rel value; <n> = 0 to 120e6 (
Ω
).
0
:RESistance:REFerence:STATe <b> [, <clist>]
Enable/disable rel; <b> = ON or OFF.
OFF
:RESistance:REFerence:ACQuire [, <clist>]
Use input signal as rel value.
Rel commands for
Ω
4
[SENSe[1]]
Optional root command.
:FRESistance:REFerence <n> [, <clist>]
Specify rel value; <n> = 0 to 120e6 (
Ω
).
0
:FRESistance:REFerence:STATe <b> [, <clist>]
Enable/disable rel; <b> = ON or OFF.
OFF
:FRESistance:REFerence:ACQuire [, <clist>]
Use input signal as rel value.
Rel commands for TEMP
2
[SENSe[1]]
Optional root command.
:TEMPerature:REFerence <n> [, <clist>]
Specify rel value; <n> = -200 to 1820 (°C). 0
:TEMPerature:REFerence:STATe <b> [, <clist>]
Enable/disable rel; <b> = ON or OFF.
OFF
:TEMPerature:REFerence:ACQuire [, <clist>]
Use input signal as rel value.
Rel commands for FREQ
[SENSe[1]]
Optional root command.
:FREQuency:REFerence <n> [, <clist>]
Specify rel value; <n> = 0 to 1.5e7 (Hz).
0
:FREQuency:REFerence:STATe <b> [, <clist>]
Enable/disable rel; <b> = ON or OFF.
OFF
:FREQuency:REFerence:ACQuire [, <clist>]
Use input signal as rel value.
Table 5-1 (continued)
Rel commands
2750-900-01.book Page 5 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...