![background image](http://html1.mh-extra.com/html/keithley/2750/2750_user-manual_3956564049.webp)
1-32
Getting Started
Model 2750 Multimeter/Switch System User’s Manual
Timer interval (TIMER) — For the STEP function, the timer specifies the time delay
between scanned channels. For the SCAN function, the interval specifies the time delay
between scans. The timer starts when the scan is started. For SCAN, the next scan will not
start until the timer interval expires.
NOTE
The Model 2750 can also be configured to run an advanced scan. For an
advanced scan, each channel can have its own unique setup (i.e., function,
range, etc.). Advanced scanning is covered in
Figure 1-8
Simple scan operation
MIN CHAN: XXX
INT: SIMPLE
Display SIMPLE option
and press ENTER
Specify minimum channel
(XXX) and press ENTER
SHIFT
SCAN
STEP
CONFIG
Press SHIFT
Press CONFIG (STEP)
MAX CHAN: YYY Specify maximum channel
(YYY) and press ENTER
TIMER? NO/YES
Display NO or YES and
press ENTER
YES
NO
xxH:xxM:xx.xxxS
Set timer interval in
hr:min:sec format and
press ENTER
RDG CT:xxxxxx
Specify reading count and press
ENTER
SCAN
STEP
Press STEP or SCAN to start
scan
Timer interval specifies time
between scans.
Reading count:
Specifies number of scans
to be performed.
Specifies number of readings
to store in buffer.
Timer interval specifies time between
scanned channels.
Reading count:
Specifies number of channels to be
scanned.
Specifies number of readings to store
in buffer.
Step 1. Configure simple scan:
Step 2. Run simple scan:
SHIFT
SCAN
STEP
HALT
Press SHIFT
Press HALT (SCAN)
Step 3. Disable scan mode:
2750-900-01.book Page 32 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...