4-6
Range, Digits, Rate, Bandwidth, and Filter
Model 2750 Multimeter/Switch System User’s Manual
Remote programming — digits
Digits commands
The commands to control display resolution (digits) are listed in
. Additional
information on these commands follow the table.
NOTE
Query commands are not included in
. All commands for the SENSe
subsystem are provided in
Table 4-3
Digits commands
Commands*
Description
Default
[SENSe[1]]
Optional root command.
:VOLTage[:DC]:DIGits <n> [, <clist>]
Set # of digits for DCV; <n> = 4 to 7.
7
:VOLTage:AC:DIGits <n> [, <clist>]
Set # of digits for ACV; <n> = 4 to 7.
6
:CURRent[:DC]:DIGits <n> [, <clist>]
Set # of digits for DC1; <n> = 4 to 7.
7
:CURRent:AC:DIGits <n> [, <clist>]
Set # of digits for ACI; <n> = 4 to 7.
6
:RESistance:DIGits <n> [, <clist>]
Set # of digits for
Ω
2; <n> = 4 to 7.
7
:FRESistance:DIGits <n> [, <clist>]
Set # of digits for
Ω
4; <n> = 4 to 7.
7
:TEMPerature:DIGits <n> [, <clist>]
Set # of digits for TEMP; <n> = 4 to 7.
6
:FREQuency:DIGits <n> [, <clist>]
Set # of digits for FREQ; <n> = 4 to 7.
7
:PERiod:DIGits <n> [, <clist>]
Set # of digits for PERIOD; <n> = 4 to 7.
7
Channel list parameter:
<clist> = (@SCH)
where: S
= Mainframe slot number (1, 2, 3, 4 or 5)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
* The <clist> parameter is used to configure one or more channels for a scan. Each channel in the <clist> must be set to the function
specified by the digits command. If not, a conflict error (-221) will occur. For example, VOLTage:AC:DIGits 4.5, (@101) is only
valid if scan channel 101 is set for the ACV function.
2750-900-01.book Page 6 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...