8-4
Triggering
Model 2750 Multimeter/Switch System User’s Manual
The delay function is accessed by pressing SHIFT and then DELAY. The present delay
setting (AUTO or MANual) is displayed. Press the
Δ
or
∇
key to display the desired set-
ting and press ENTER.
If MANual is chosen, also enter the duration of the delay in the hour/minute/second for-
mat using the
,
,
Δ
, and
∇
keys. The maximum is 99H:99M:99SS. Note that press-
ing the AUTO key sets the delay to 0.001 sec. Press ENTER to accept the delay or EXIT
for no change.
Table 8-1
Auto delay settings
Function
Range and delay
DCV
100mV
1ms
1V
1ms
10V
1ms
100V
5ms
1000V
5ms
ACV
100mV
400ms
1V
400ms
10V
400ms
100V
400ms
750V
400ms
FREQ and
PERIOD
100mV
1ms
1V
1ms
10V
1ms
100V
1ms
750V
1ms
DCI
20mA
2ms
100mA
2ms
1A
2ms
3A
2ms
ACI
1A
400ms
3A
400ms
Ω
2,
Ω
4
1
1
Ω
2
3ms
10
Ω
3ms
100
Ω
3ms
1k
Ω
3ms
10k
Ω
13ms
100k
Ω
25ms
1M
Ω
100ms
10M
Ω
150ms
100M
Ω
250ms
Dry Circuit
Ohms
3
1
Ω
3ms
10
Ω
3ms
100
Ω
13ms
1k
Ω
13ms
Continuity
1k
Ω
3ms
TEMP
The auto delay for thermocouples is 1ms. For thermistors and 4-wire RTDs, the auto delay
period is the same as the delay for the resistance range that is used for the measurement.
1. Delay times also apply for offset-compensated ohms (OCOMP).
2. 1
Ω
range not available for the
Ω
2 function.
3. State (on or off) of offset-compensated ohms (OCOMP) does not affect delay times.
2750-900-01.book Page 4 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...