7-38
Scanning
Model 2750 Multimeter/Switch System User’s Manual
Table 7-3
Monitor scan example
Front panel operation
Remote programming
1
Restore defaults (SHIFT SETUP > RESTORE: FACT).
SYST:PRES
2
For front panel operation, proceed to step 3.
For remote programming, clear the buffer:
TRAC:CLE
3
Configure advanced scan:
SHIFT CONFIG > ADVANCED:
a
Channel 101:
Select TEMP function.
FUNC 'TEMP',(@101)
Configure temperature (SHIFT SENSOR):
Select Thermocouple sensor (SENS: TCOUPLE).
TEMP:TRAN TC,(@101)
Select type K thermocouple (TYPE: K).
TEMP:TC:TYPE K,(@101)
Select internal reference junction (JUNC: INT).
TEMP:RJUN:RSEL INT,(@101)
b
Set and enable high limit 1:
Set limit to 30 (SHIFT LIMITS > HI1:+30.00000).
CALC3:LIM1:UPP 30,(@101)
Enable (on) limit (SHIFT OFF/ON > LIMITS: ON).
CALC3:LIM1:STAT ON,(@101)
Enable Channel Average (SHIFT CH AVG).
CAV ON,(@101)
c
Channel 102, 103, and 104:
Select DCV function.
FUNC 'VOLT',(@102:104)
Select 10V range.
VOLT:RANG 10,(@102:104)
Set filter count to 20 (SHIFT TYPE > 020 RDGS).
VOLT:AVER:COUN 20,(@102:104)
Enable filter (FILTER).
VOLT:AVER:STAT ON,(@102:104)
d
Disable (off) channels 105 through 222 (SHIFT CH OFF).
ROUT:SCAN (@101:104)
e
Disable immediate scan (IMM SCAN: N), and enable
high limit 1 (HLIM1 SCAN:Y).
ROUT:SCAN:TSO HLIM1
f
Disable timer (TIMER? OFF).
g
Set reading count to 4.
ROUT:MON:POIN 4
4
Select and enable monitor channel
(SHIFT MONITOR >101).
ROUT:MON (@101)
ROUT:MON:STAT ON
2750-900-01.book Page 38 Wednesday, August 3, 2011 7:56 AM
Содержание 2750
Страница 3: ...2750 900 01 book Page 2 Wednesday August 3 2011 7 56 AM...
Страница 7: ...2750 900 01 book Page 4 Wednesday August 3 2011 7 56 AM...
Страница 17: ...2750 900 01 book Page 10 Wednesday August 3 2011 7 56 AM...
Страница 204: ...Model 2750 Multimeter Switch System User s Manual Buffer 6 15 2750 900 01 book Page 15 Wednesday August 3 2011 7 56 AM...
Страница 205: ...6 16 Buffer Model 2750 Multimeter Switch System User s Manual 2750 900 01 book Page 16 Wednesday August 3 2011 7 56 AM...
Страница 324: ...12 CommonCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 332: ...13 SCPISignalOriented MeasurementCommands 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 348: ...15 SCPIReferenceTables 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 374: ...B Model7700ConnectionGuide 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 386: ...C StatusandErrorMessages 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 394: ...D SignalProcessing SequenceandDataFlow 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 408: ...E MeasurementConsiderations 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...
Страница 428: ...G IEEE 488BusOverview 2750 900 01 book Page 1 Wednesday August 3 2011 7 56 AM...