Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.5 Adjusting the patient chair
50 / 178
Motorised adjustment of the patient unit (E80 Vision
standard, E70 Vision optional)
When automatic positions "AP 0", "AP 1", "AP 2" or rinsing position "SP" of the
patient chair is saved, the position of the patient part is saved also.
There are two memory positions available:
1. Save by pressing the "SP" button:
The patient part moves into the rinse position after the chair has stopped
moving.
2. Save by pressing the buttons "AP 0", "AP 1" or "AP 2":
The patient's unit swings back into resting position.
4.5 Adjusting the patient chair
CAUTION
Danger of injury from automatic chair adjustment
Injury can arise from the automatic adjustment of the chair position.
▶
Only use the automatic functions under the supervision of the user.
Note
The lift and backrest motors have thermal fuses. The motors shut down at an
operating temperature of 105
o
C. The cooling phase lasts approx. 15
minutes. After the cooling phase is over, the lift and backrest motors are op-
erable. Such temperatures are not reached in normal practice. The shutoff
temperature can be reached when the motors are frequently actuated in
presentations and events (approximately 8 complete movement cycles).
CAUTION
The patient chair is overloaded
Damage to the support system or patient chair.
Injury to the patient or office staff.
▶
Do not exceed a maximum permissible weight of 180 kg.
▶
Do not sit on the head or foot end of the patient chair when it is horizontally
aligned.
▶
Monitor the patient when changing the chair position.
NOTICE
Direct contact between the skin of the patient and the cushion during
patient positioning
To improve hygiene and increase the service life, KaVo recommends to use ap-
propriate covers (e.g. head protection bags) to protect the cushions that are
regularly in contact with the skin of the patient. Sweat or hair and skin care
products can be a serious strain on the cushions. The cover can prevent aller-
gic reactions or skin irritations in patients with sensitive skin. Please comply
with the instructions of the manufacturer of the cover.
4.5.1 Swivelling the arm rest
The right arm rest can be swivelled forward for the patient to get in and out.
▶
Swivel arm rest forward.