Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.15 Using the SL600 surgical motor (optional accessory)
Setting the torque
Note
The torque values are only for KaVo handpieces and angle pieces that oper-
ate properly.
▶
Tap Current torque (value in Ncm).
ð
Font colour switches to blue.
▶
Use slider to set the torque.
ð
The torque is shown on the display and is effective immediately.
Setting the speed
▶
Tap the current speed (value in rpm).
ð
The font colour switches to blue.
▶
Use slider to set the speed.
ð
The speed is shown on the display and is effective immediately.
Setting the coolant
▶
Tap the „NaCl“ key to select the coolant flow rate.
▶
Press the "Preselect spray" foot button to turn the coolant on or off and to
adjust the coolant flow.
Setting the transmission factor
▶
Unfold the "Transmission ratio" list and select the transmission ratio as de-
sired. The transmission ratio cannot be saved.
Setting the direction of motor rotation
▶
Tap the "Direction of motor rotation" key to toggle between clockwise and
counterclockwise rotation.
or
▶
Slide the cross switch upward.
ð
The rotational direction of the motor is switched back and forth each time
the cross-switch is pushed: counterclockwise rotation - clockwise rotation.
ð
A signal tone sounds when the setting "Motor left rotation" is selected.
Setting activity
▶
Tap the activity symbol to change the activity of the respective step (toggle
function).