Instructions for use ESTETICA E70 Vision / E80 Vision
9 Eliminating disturbances
Malfunction
Cause
Remedy
Camera image freezes
without the release
button or foot control
having been triggered.
Taking the camera out
again did not solve the
problem.
Electrical or electro-
magnetic interference
by other equipment.
▶
Restart the software.
Camera image freezes
without the release
button or foot control
having been triggered.
The monitor turns itself
off.
Electrical or electro-
magnetic interference
by other equipment.
▶
Restart the treatment unit and the CONEXIO
PC.
Additional warning messages
Malfunction
Cause
Remedy
A signal is emitted
every 10 seconds and a
status message is
shown.
The oxygenal container
is empty.
▶
Refill the Oxygenal container.
Ten beeps are issued.
The Oxygenal container
is too full.
▶
Stop filling the Oxygenal container.
A melody sounds.
The amalgam separator
CAS1 is almost full
(95%).
▶
Exchange the amalgam container.
The CAS1 amalgam
separator is defective.
▶
Also refer to:
Instructions for use for the CAS
1 or
▶
Call a Service technician.
An acoustic signal is is-
sued every second.
Leaking water switch
recognises leaking wa-
ter.
▶
Remove water from the unit body. If neces-
sary, have a technician fixed the leak.
Error messages on the touchscreen
Malfunction
Cause
Remedy
Status message ""Unit
controller not recog-
nised""
Cable or electronics is-
sue.
▶
Call a service technician.