Instructions for use ESTETICA E70 Vision / E80 Vision
7 Safety check - Test instructions | 7.2 Instructions for the safety check
Using the measuring cable
①
the unit is disconnected from the mains supply
and connection of the treatment centre to the safety tester is enabled. Hence,
the customer-provided mains supply L & N on the power input board need not
be disconnected. The adapter cable
②
is included in the delivery of the KaVo
measuring cable and is required for older treatment centres that are not
equipped with an X2 connector.
Connecting the safety tester to KaVo measuring cable
on the treatment centre
▶
Remove plug X2 from the power input board and plug it into the matching
connector X2 of the KaVo- measuring cable (
Mat. no. 0.411.8811
).
▶
Plug the second plug X2 of the KaVo measuring cable into the network card
(X2).
▶
Insert the protective contact plug of the KaVo measuring cable into the
safety tester.
Connecting the safety tester to the treatment centre
without a KaVo measuring cable
▶
Switch L + N of the on-site power supply cord to be voltage-free.
▶
Disconnect L + N on the mains terminals.
▶
Connect the safety tester directly to terminals mains and and protective
earth conductor terminal X9.
155 / 178