Instructions for use ESTETICA E70 Vision / E80 Vision
3 Product description | 3.10 Technical data
Triple-function handpiece and multifunctional handpiece
Water pressure
1.5 ± 0.3 bar; Flow pressure; 4 x
manometer
Max. static pressure water
2.5 ± 0.3 bar
Water flow
80 ± 10 ml/min
Air pressure
3.3 ± 0.1 bar; Flow pressure; 4 x
manometer
Max. static pressure air
4 + 0.5 bar
Air flow
at least 16 Nl/min
Operating time (multifunctional
handpiece only)
1 minute
Interval (multifunctional handpiece
only)
3 minutes
Electrical multifunctional handpiece
Safety extra-low voltage according
to DIN EN 60601-1:
24 V AC ± 10% (non-grounded
voltage)
Frequency
50/60 Hz
Type of use
B
Heat output for water
approx. 90 W
Heat output for air
approx. 20 W
Lamp voltage
max. 3.2 V ± 0.15 V
High-pressure lamp power
max. 2.5 W
Water supply
Note
If the water is very hard (above 12 °dH), a water softening device must be
fitted in the ion-exchange process.
Insufficient water hardness (below 8.4 °dH) can promote the formation of al-
gae.
Note
In conjunction with the "DVGW water block with integrated water germ re-
duction" a water germ reduction unit is installed in dental units from KaVo.
The germ reduction liquid OXYGENAL 6 cleaner is continually added to the
water at a concentration which is harmless for persons, but hygienically ef-
fective to maintain the quality of the treatment water by making it more diffi-
cult for microorganisms to multiply and spread. The handling is described in
the care instructions of the treatment centres. Supplementary measures such
as the rinsing of water conducting lines and intensive germ reduction must
be carried out according to the instructions of the manufacturer.