Instructions for use ESTETICA E70 Vision / E80 Vision
7 Safety check - Test instructions | 7.2 Instructions for the safety check
1
1
▶
Connect TP2 measuring socket
①
to the safety tester.
1
▶
Connect
①
ultrasonic scaler tip to the safety tester.
Note
Additional measuring points AP X must be taken into consideration in the
presence of accessories: e.g. accessories such as PIEZO ultrasound scaler.
See also:
2
8 Appendix - Additional protective conductor measuring sites, Page 167
Connecting accessible conductive parts [ACP] to PE
ACP = accessible conductive parts
157 / 178