Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.10 Using functions through the touchscreen
Symbol
Function
No cooling
Spray water cooling
Dosing the amount of spray water
CAUTION
Lack of working tip cooling.
Heat damage to tooth or handpiece.
▶
Never work under dry conditions, except in case of tips designed for these
conditions.
▶
Set a minimum flow rate of 6 ml/min. For this purpose, adjust the flow rate
such the water is just between dripping and flowing during irrigation.
▶
For the amount of spray water for each tip, please refer to the PiezoLED In-
structions for Use.
See also:
2
Instructions for Use PiezoLED
▶
Adjust the amount of spray water using the regulating ring.
▶
After you set a single value or all values, tap the "Save" key to save the val-
ues.
Settings for the multifunctional handpiece
The following settings can be changed in the Treatment menu on the touch-
screen:
▪ Air/water heating
▶
Take the multifunctional handpiece off the holder.
69 / 178