Instructions for use ESTETICA E70 Vision / E80 Vision
9 Eliminating disturbances
170 / 178
Malfunction
Cause
Remedy
The patient chair does
not move.
The safety shutoff is
activated.
Foot control is activ-
ated.
▶
Check the safety shutoffs and eliminate the
reason for the shutoff.
Display without indic-
ator.
Bus / hardware error.
▶
Turn the device off and on.
▶
Call the service technician to look into the
problem if it continues to exist.
Operating device no
function.
Bus / hardware error.
▶
Turn the device off and on.
▶
Call the service technician to look into the
problem if it continues to exist.
Several handpieces are
simultaneously activ-
ated.
Hardware error.
▶
Stop working and call the service technician
The treatment unit is
not connected to the
wireless foot control.
The treatment unit
emits a tone.
The wireless foot con-
trol is turned off.
▶
Check the on/off switch on the foot control and
turn it on if necessary.
The wireless foot con-
trol is out of range.
▶
Move the wireless foot control into the range of
the treatment unit.
Malfunction or low bat-
tery
▶
Check the status display of the foot control.
Yellow: low battery
No display: malfunction
▶
Charge battery.
The wireless foot con-
trol does not return to
centre position.
▶
Charge battery.
Malfunction
Cause
Remedy
Camera images shows
images only as black/
white images.
Electrical or electro-
magnetic interference
by other equipment.
▶
Restart the CONEXIO PC.
Camera image freezes
without the release
button or foot control
having been triggered.
Camera image fails to
return to live image
mode.
Electrical or electro-
magnetic interference
by other equipment.
▶
Replace the camera in the holder and then
take it out again.