Instructions for use ESTETICA E70 Vision / E80 Vision
3 Product description | 3.10 Technical data
Air inlet pressure
5.2 to 7 bar (0.52 to 0.7 MPa)
Minimum air flow rate
min. 80 Nl/min
Pressure dew point
< -30 ºC (compressor with
dry air system)
Oil content
< 0.1 mg/m
3
(oil-free compressor)
Contamination
< 100 particles/m
3
at particle sizes
of 1 to 5 µm
Air connection
R 1/2"
Air connection above floor level
min. 50 mm, max. 105 mm with
valve opened
Suction
Pressure drop at the connecting point
①
Semi-dry and wet suction
②
Dry suction
Note
If the negative static pressure is > 180 mbar, the unit must be equipped with
the negative pressure regulating valve assembly kit.
Cannula connecting pieces
according to DIN EN ISO 7494 - 2
Diameter of cannula of saliva ejector
and surgical suction
7 mm
Diameter of cannula of spray mist
suction
15 mm
Suction air quantity at spray mist
cannula
minimum 250Nl/min (suction system
with high flow rate), recommended
300Nl/min
Diameter of the suction connection
40 mm
Suction connection above floor
20 mm