4 Operation | 4.16 Use pump for physiological saline solution (optional accessory)
134 / 178
4.16.1 General information
CAUTION
Running, open hose pump.
Risk of injury.
▶
Turn off the device before opening the hose pump.
CAUTION
Coolant container made of glass.
Danger of injury due to falling or shattering glass coolant container.
▶
Do not use glass bottles as coolant containers on the dentist part of the
dental unit.
CAUTION
Danger of tipping due to the coolant containers being too heavy.
Malfunctions.
▶
Use coolant containers with a maximal volume of 1 litre only.
▶
Check the stability.
CAUTION
Use of non-sterile coolant hose with accessory.
Infection hazard.
▶
The sterile hose kit is only designed for single use, do not use again.
▶
Dispose of the sterile hose kit in the prescribed manner.
▶
Use a new, sterile packed coolant hose with accessory for every treatment.
Note
All liquid-conducting parts are not sterile. They must be sterilised before the
first treatment. All parts conducting liquids must be kept sterile.
Note
The coolant must be selected to suit the planned application. The flow rate of
the coolant is dependent on the instrument used. The user must set an ad-
equate flow of coolant and check this.
Note
Check the integrity of the hose set before use. If product or packaging are
damaged, the product needs to be discarded.
Note
The correct flow direction must be observed when the hose is inserted into
the pump.
The physiological saline solution may only be used in conjunction with NaCl-
resistant instruments.
The following symbols are displayed on the sterile hose kit S 600 (10 pcs)
(
Mat. no. 1.009.8757
):
Icon
Description
Date of manufacture