Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.19 Using the camera
142 / 178
CAUTION
Electrical power.
Electrical shock from incorrectly connecting a non-medical system to the USB
interfaces of the device.
▶
Connect any IT device to the medical system in accordance with IEC
60601-1.
▶
Use USB devices with no additional power supply (USB-powered) only.
▶
Applied parts connected to the USB interface of the dentist element must
comply with the requisite insulation.
▶
USB-powered devices failing to meet the requisite insulation for applied
parts must be placed appropriately such that direct contact of the USB
device and the patient is excluded.
▶
It is not permissible to touch USB-powered devices failing to meet the re-
quisite insulation for applied parts and the patient at the same time.
The treatment unit may be fitted with up to three USB ports. Camera interfaces
are situated on the underside of the dentist element (T-table) or in the dentist
element (S-table). Only the cameras approved/enclosed in the delivery by KaVo
may be connected to these interfaces.
The USB port in the back is connected directly to the back-of-the-head PC (in
the presence of the corresponding wiring). USB devices meeting the specifica-
tions listed above can be connected to this interface. To use USB devices that
have been connected, it may be necessary to install a suitable driver software
on the back-of-the-head PC.
Getting the USB ports ready for use
▶
To run an USB device, connect the USB port in the terminal box of the
treatment centre to an external back-of-the-head PC. Use one or maximally
two USB extension cables 5 m (
Mat. no. 1.004.6953
) according to need.
▶
USB devices connected to the dentist element must meet the USB stand-
ards, USB 1.0, 1.1 or 2.0, and consume max. 500 mA of electrical power.
4.19 Using the camera
See also:
2
Instructions for Use ERGOcam One
See also:
2
Instructions for Use DIAGNOcam 2170 U
4.20 Service table 1568 (optional accessory)
NOTICE
Exceeding the load limits.
Damage to the service table.
▶
Comply with maximal load limits.
NOTICE
Over-travel beyond swivel range.
Property damage.
▶
Comply with rotary knob swivel range of 180°.