Instructions for use ESTETICA E70 Vision / E80 Vision
7 Safety check - Test instructions | 7.2 Instructions for the safety check
Protective earth measurement
The protective conductor resistance must be measured at the following parts of
the device:
▪ Treatment centre
▪ Treatment lamp
▪ Accessories
Note
Additional measuring points SL X need to be taken into consideration in the
presence of accessories: e.g. accessories such as connection to external
devices, camera module of the patient communication system, etc.
See also:
2
8 Appendix - Additional protective conductor measuring sites, Page 167
Scanning the treatment centre with the test tip
Body base
①
Surroundings of the protective
earth conductor terminal
②
Main switch plate
③
Switching power supply
159 / 178