Instructions for use ESTETICA E70 Vision / E80 Vision
4 Operation | 4.10 Using functions through the touchscreen
62 / 178
CAUTION
Danger of crushing when retracting the patient chair with activated
safety shutoff.
The patient can be clamped in.
▶
Only retract the patient chair without the patient.
To allow the chair to move freely, it can also be moved when the safety circuit
is on. Use this function for repair purposes only.
▶
Press and hold down the "SP" and "LP" keys simultaneously on the assistant
element and the foot control.
▶
Move the chair using the button wheel buttons of the chair.
CAUTION
Safety shutoff disabled, move motors without monitoring.
Destruction of the motor.
▶
Monitor the path of travel of the motor.
▶
Do not move the motor to the block.
▶
For all chair movements, remove all restrictions from the swinging range of
the chair.
4.10 Using functions through the touchscreen
The use of the touchscreen is simple and always follows the same pattern.
The touchscreen is subdivided into five menus (tabs):
▪ Treatment menu
▪ Timer menu
▪ CONEXIO menu
▪ Cleaning menu
▪ Settings menu
▶
Tap a tab to display the corresponding menu.
Tab
Menu
Description
Treatment
▪ Selection of type of treatment
▪ Speed / power
▪ Direction of motor rotation, if
applicable
▪ Cooling status