Instructions for use ESTETICA E70 Vision / E80 Vision
3 Product description | 3.10 Technical data
40 / 178
WARNING
There is a risk of infection if you fail to comply with national regula-
tions.
Contamination of the treatment water and/or drinking water supply with
germs.
▶
Note and comply with national regulations concerning the quality of water
for human use (drinking water) - if applicable.
▶
Note and comply with national regulations concerning the prevention of
backflow (flow from the treatment centre back to the public water supply) -
if applicable.
According to DIN EN 1717, each unit that is not listed by DVGW must be
provided with an upstream type AA, AB or AD safety device.
When establishing a water connection, prevent brackish water pools with stand-
ing water (also in the house plumbing).
For further information, please refer to www.dvgw.de
Free drainage according to DIN EN
1717 - DVGW certified
Register No.: AS-0630BT0111
Water quality
Tap water, cold water connection
Water hardness
1.5 to 2.14 mmol/l
≙
8.4 to 12 °dH
pH
7.2 to 7.8
Customer-site water filtering
80 microns
Water connection
Shut-off valve with brass cone com-
pression screw connection 3/8" to Ø
10 mm provided
Above-floor water connection
min. 50 mm, max. 105 mm with
valve opened
Water inlet pressure
2.0 to 6.0 bar (0.2 to 0.6 MPa)
Water inflow
4 l/min
Diameter of the drain connection
40 mm
Above-floor drain connection
20 mm
Outflow quantity
max. 4 l/min
Slope of water drain pipe
downstream from device: at least 10
mm per metre
Air supply
WARNING
Non-adherence to national guidelines regarding the quality of the
dental air.
Infection hazard.
▶
Observe and adhere to the national guidelines regarding the quality of the
dental air - if any.
▶
Blow through the air line prior to commissioning.