8 Diagnosis / error messages
73
Error no. /
Area
Text on the LCD display
Current output
Digital output
Pulse output
LCD display
Maskable error?
C078.004 / Config Flowrate to zero.
Check digital in terminals.
4 mA (0 % flow)
No change; current
value.
0
0 %
“...Group Masking”
menu.
C076.006 / Config All totalizer stopp.
Check digital in terminals.
No change; current
value.
No change; current
value.
No change; current
value.
No change; current
value.
C074.007 / Config Totalizer reset.
Reset of one or more Totalizers.
No change; current
value.
No change; current
value.
No change; current
value.
No change; current
value.
“...Group Masking”
menu.
C072.003 / Config Simulation is on Simulated
values.
Switch off Simulation Mode.
No change; current
value.
No change; current
value.
No change; current
value.
No change; current
value.
C070.027 / Config An alarm is simulated.
Switch off alarm simulation.
No change; current
value.
No change; current
value.
No change; current
value.
No change; current
value.
S065.044 /
Operation
Inhouse temp. alarm.
Reduce ambient temperature.
No change; current
value.
No response
No response
No change; current
value.
S064.041 /
Operation
EPD alarm.
Secure pipe is completely filled.
Alarm – as configured
in the “Curr. at EPD
Alarm” menu.
Alarm, if the digital
output has been
configured as a “Logic
/ Alarm Signal / EPD”.
0
0 %
“...Group Masking”
menu.
S062.039 /
Operation
Sensor temp. limits alarm.
Change limits or change fluid
temperature.
No change; current
value.
No response
No change; current
value.
No change; current
value.
“...Group Masking”
menu.
S052.017 /
Operation
Curr.Out 31 / 32 is saturated.
CO process value out of range.
Adapt Qmax.
Alarm – as configured
in the “Curr.Out >
20.5mA” menu.
No change; current
value.
No change; current
value.
No change; current
value.
“...Group Masking”
menu.
S047.016 /
Operation
Pulse output is cut off.
Wrong config. Check pulse out
configuration.
No change; current
value.
No change; current
value.
Maximum pulse rate
possible
No change; current
value.
“...Group Masking”
menu.
S046.001 /
Operation
Mass flowrate exceeds limits.
Check flowrate and alarm limits.
No change; current
value.
Alarm, if the digital
output has been
configured as a “Qm
Massflow Max” or a
“Qm Massflow Min”.
No change; current
value.
No change; current
value.
“...Group Masking”
menu.
S044.002 /
Operation
Volume flowrate exceeds limits.
Check flowrate and alarm limits.
No change; current
value.
Alarm, if the digital
output has been
configured as a “Qv
Volumeflow Max” or a
“Qv Volumeflow Min”.
No change; current
value.
No change; current
value.
“...Group Masking”
menu.
S041.034 /
Electronics
DC feedback regulation.
Check conditions of application.
Call service.
4 mA (0 % flow)
No response
0 Hz
0 %
“...Group Masking”
menu.
S040.031 /
Electronics
Coil impedance alarm.
Call service.
No change; current
value.
No response
No change; current
value.
No change; current
value.
“...Group Masking”
menu.
Содержание 406060
Страница 2: ......
Страница 98: ...13 Appendix 98 13 3 Declarations of conformity 13 3 1 JUMO flowTRANS MAG S10...
Страница 99: ...13 Appendix 99...
Страница 100: ...13 Appendix 100...
Страница 101: ...13 Appendix 101...
Страница 102: ...13 Appendix 102...
Страница 103: ...13 Appendix 103...
Страница 104: ...13 Appendix 104 13 3 2 JUMO flowTRANS MAG H10...
Страница 105: ...13 Appendix 105...
Страница 106: ...13 Appendix 106...
Страница 107: ...13 Appendix 107...
Страница 108: ...13 Appendix 108...
Страница 109: ...13 Appendix 109...
Страница 110: ...13 Appendix 110 13 3 3 JUMO flowTRANS MAG 10...
Страница 111: ...13 Appendix 111...
Страница 112: ...13 Appendix 112...
Страница 113: ...13 Appendix 113...
Страница 115: ......