6 Commissioning
37
6.3
Checks prior to commissioning
The following points must be checked before commissioning the
device:
• Correct wiring in accordance with
Electrical connections
on page 26.
• Correct grounding of the sensor.
• The ambient conditions must meet the requirements set
out in the specification.
• The power supply must meet the requirements set out on
the name plate.
Remote mount design – check sensor and
transmitter for correct assignment
Figure 45: Assignment of sensor to transmitter
The correct assignment between sensors and transmitters has to
be ensured for devices featuring a remote mount design. Devices
that form a pair have the same numeric suffix and this is indicated
on the rating plate; e. g., the X001 sensor is paired with the Y001
transmitter, and the X002 sensor is paired with the Y002
transmitter.
1
SensorMemory
Figure 46: SensorMemory
The SensorMemory is a removable memory module located at
the rear of the transmitter electronics module.
The part number and the numeric suffix are indicated on the
SensorMemory module.
The numeric suffix is also indicated on the rating plate of the
corresponding sensor.
The numeric suffix for the SensorMemory module has to be
consistent with the numeric suffix for the sensor.
6.4
Parameterization of the device
Commissioning and operation of the JUMO flowTRANS MAG
S10/H10 can be carried out via the integrated LCD display
(optional; see
Parameterization via the menu function Easy
Setup
on page 38).
Parameter settings using the optional LCD display
1
Connector for LCD display
2
LCD display
Figure 47: Optional LCD display
Note
For devices without an LCD display, an accessory LCD display is
available that can be connected (part number: 00706388) to
make the parameter settings.
Содержание 406060
Страница 2: ......
Страница 98: ...13 Appendix 98 13 3 Declarations of conformity 13 3 1 JUMO flowTRANS MAG S10...
Страница 99: ...13 Appendix 99...
Страница 100: ...13 Appendix 100...
Страница 101: ...13 Appendix 101...
Страница 102: ...13 Appendix 102...
Страница 103: ...13 Appendix 103...
Страница 104: ...13 Appendix 104 13 3 2 JUMO flowTRANS MAG H10...
Страница 105: ...13 Appendix 105...
Страница 106: ...13 Appendix 106...
Страница 107: ...13 Appendix 107...
Страница 108: ...13 Appendix 108...
Страница 109: ...13 Appendix 109...
Страница 110: ...13 Appendix 110 13 3 3 JUMO flowTRANS MAG 10...
Страница 111: ...13 Appendix 111...
Страница 112: ...13 Appendix 112...
Страница 113: ...13 Appendix 113...
Страница 115: ......