8 Diagnosis / error messages
69
8.2 Error
messages
The error messages are divided into four groups in accordance with the NAMUR classification scheme.
Errors
Error no. / Area
Text on the LCD display
Cause
Remedy
F099.042 / Electronics
ADC overrange alarm.
Noise too high. Check applicati.
Call service.
The signal at the input of the A / D converter
exceeds the maximum value.
Flow measurement is not possible.
When the meter tube is empty, ensure that the
empty pipe detection function is active.
Ensure that the current flow rate does not exceed
the set measuring range.
Contact the manufacturer.
F098.012 / Electronics
FEB communication error.
EMC disturbance. Call Service.
Defective front end board or motherboard.
Wiring error between sensor and transmitter.
The front end board and the motherboard are not
connected.
If this involves an integral mount design: Contact
the manufacturer.
If this involves a remote mount design: Check the
wiring between the sensor and the transmitter.
Contact the manufacturer.
F097.029 / Electronics
Coil regulation error.
Check wiring of sensor coils. Call
service
Defective solenoid.
Wiring error, broken cable or shorted out coil circuit
(terminals M1 / M2).
Defective fuse for coil circuit; moisture in sensor
terminal box.
Check wiring.
Check coil circuit fuse.
Check the sensor terminal box for moisture.
F096.043 / Electronics
SIL self check alarm.
Call service.
The SIL monitoring function has detected a
transmitter error.
Contact the manufacturer.
F095.036 / Electronics
Coil isolation alarm.
Call service.
Defective solenoid.
Wiring error with coil circuit (short-circuit, terminals
M1 / M2 downstream of GND).
Flooded sensor.
Contact the manufacturer.
F094.022 / Electronics
Safety Alarm Curr. Out 31 / 32
SIL function detects error. Call
Service.
SIL-relevant error detected at current output.
Contact the manufacturer.
F093.032 / Electronics
Electrode short cuircit.
Check wiring of sensor electrode.
Call service.
Wiring error with electrode circuit (short-circuit
downstream of shield).
Leaking electrode.
Flooded sensor.
Check wiring.
Check electrode wiring for short-circuit to shield.
Check electrode for leakage.
Check the sensor terminal box for moisture.
F092.033 / Electronics
Electrode open cuircit.
Check wiring of sensor electrode.
Call service.
Wiring error with electrode circuit (interruption,
broken cable).
Check wiring.
F091.030 / Electronics
Coil wiring error.
Check wiring of sensor coils. Call
service.
Defective solenoid.
Wiring error, broken cable or shorted out coil circuit
(terminals M1 / M2).
Defective fuse for coil circuit; moisture in sensor
terminal box.
Check wiring.
Check coil circuit fuse.
Check the sensor terminal box for moisture.
F090.035 / Electronics
ADC RX210 com. error.
Call service.
Problems with the electromagnetic compatibility
(EMC) in the vicinity.
Error with transmitter electronics.
Replace the transmitter or contact the
manufacturer.
F088.013 / Electronics
FEB communication error.
EMC disturbance. Call Service.
Problems with the electromagnetic compatibility
(EMC) of the signal cable.
Incorrect signal cable.
Check the signal cable and the wiring.
Contact the manufacturer.
Содержание 406060
Страница 2: ......
Страница 98: ...13 Appendix 98 13 3 Declarations of conformity 13 3 1 JUMO flowTRANS MAG S10...
Страница 99: ...13 Appendix 99...
Страница 100: ...13 Appendix 100...
Страница 101: ...13 Appendix 101...
Страница 102: ...13 Appendix 102...
Страница 103: ...13 Appendix 103...
Страница 104: ...13 Appendix 104 13 3 2 JUMO flowTRANS MAG H10...
Страница 105: ...13 Appendix 105...
Страница 106: ...13 Appendix 106...
Страница 107: ...13 Appendix 107...
Страница 108: ...13 Appendix 108...
Страница 109: ...13 Appendix 109...
Страница 110: ...13 Appendix 110 13 3 3 JUMO flowTRANS MAG 10...
Страница 111: ...13 Appendix 111...
Страница 112: ...13 Appendix 112...
Страница 113: ...13 Appendix 113...
Страница 115: ......