8 Diagnosis / error messages
70
Error no. / Area
Text on the LCD display
Cause
Remedy
F086.019 / Electronics
Curr.Out 31 / 32 com error.
Defective Board. EMC
disturbance. Call Service.
Error with electronics involving motherboard and
transmitter.
Problems with the EMC.
Contact the manufacturer.
F084.011 / Electronics
NV data defect.
Data storage irreparable. Call
Service.
Data error in SensorMemory.
Contact the manufacturer.
F082.014 / Electronics
Incompatible Frontend Board.
Frontend not fit to Motherboard.
Call Service.
Incompatible front end board.
Contact the manufacturer.
F081.026 / Electronics
MB voltages outside range.
Defective Motherboard HW. Call
Service.
Defective motherboard of transmitter.
Replace the transmitter or contact the
manufacturer.
Function check
Error no. / Area
Text on the LCD display
Cause
Remedy
C078.004 / Config.
Flowrate to zero.
Check digital in terminals.
External deactivation is active via digital input.
Check status of the digital input.
Check parameter settings.
C076.006 / Config.
All totalizer stopp.
Check digital in terminals.
External meter stop function is active via digital
input.
Check status of the digital input.
Check parameter settings.
C074.007 / Config.
Totalizer reset.
Reset of one or more Totalizers.
External meter reset function is active via digital
input.
Check status of the digital input.
Check parameter settings.
C072.003 / Config.
Simulation is on Simulated
values.
Switch off Simulation Mode.
The simulation mode is active.
Deactivate the simulation mode.
C070.027 / Config.
An alarm is simulated.
Switch off alarm simulation.
The alarm simulation function is active.
Deactivate the alarm simulation functions.
Operation outside of the specifications (Out Of Specs)
Error no. / Area
Text on the LCD display
Cause
Remedy
S065.044 / Operation
Inhouse temp. alarm.
Reduce ambient temperature.
The ambient temperature or the temperature of the
medium being measured is outside of the
permissible range.
Check the ambient temperature and the
temperature of the medium being measured;
reduce the temperature.
S064.041 / Operation
EPD alarm.
Secure pipe is completely filled.
The meter tube is not completely full.
Ensure that the meter tube is always filled all the
way.
S062.039 / Operation
Sensor temp. limits alarm.
Change limits or change fluid
temperature.
The temperature of the medium being measured is
outside of the permissible range.
Check the temperature of the medium being
measured; adjust the alarm limits.
S052.017 / Operation
Curr.Out 31 / 32 is saturated.
CO process value out of range.
Adapt Qmax.
Current output 31 / 32, over-traveled. The selected
process value exceeds the set measuring range.
Check the setting for the measuring range and
adjust as needed.
S047.016 / Operation
Pulse output is cut off.
Wrong config. Check pulse out
configuration.
The pulse rate or the frequency at the pulse output
is outside of the permissible limits.
Check the configuration of the pulse output
parameters.
Содержание 406060
Страница 2: ......
Страница 98: ...13 Appendix 98 13 3 Declarations of conformity 13 3 1 JUMO flowTRANS MAG S10...
Страница 99: ...13 Appendix 99...
Страница 100: ...13 Appendix 100...
Страница 101: ...13 Appendix 101...
Страница 102: ...13 Appendix 102...
Страница 103: ...13 Appendix 103...
Страница 104: ...13 Appendix 104 13 3 2 JUMO flowTRANS MAG H10...
Страница 105: ...13 Appendix 105...
Страница 106: ...13 Appendix 106...
Страница 107: ...13 Appendix 107...
Страница 108: ...13 Appendix 108...
Страница 109: ...13 Appendix 109...
Страница 110: ...13 Appendix 110 13 3 3 JUMO flowTRANS MAG 10...
Страница 111: ...13 Appendix 111...
Страница 112: ...13 Appendix 112...
Страница 113: ...13 Appendix 113...
Страница 115: ......