7 Operation
57
7.5.4 Menu: Device Setup
Menu / Parameters
Description
Device Setup
...Access Control
Used to select the “...Access Control” submenu with
.
...Sensor
Used to select the “...Sensor” submenu with
.
...Transmitter
Used to select the “...Transmitter” submenu with
.
Device Setup / ...Access Control
Standard Password
Used to enter / change the password for the “Standard” access level.
Read Only Switch
Used to indicate the position of the write-protection switch.
Further information can be found in the chapter on
Write-protection switch, service LED and local operator interface
on
page 36.
Device Setup / ...Sensor
Range Mode Config
Used to activate the second measuring range for mass flow and volumetric flow.
The setting can be made separately for the mass flow (Qm) and the volumetric flow (Qv). This enables you to quickly toggle
between two measuring ranges (e. g. Qm max and Qm max2). The change-over occurs via the “Qm Range Mode” and “Qv
Range Mode” parameters.
• Disabled: Used to deactivate the second measuring range for mass flow and volumetric flow.
• Qm and Qv: Used to activate the second measuring range for mass flow and volumetric flow.
• Qm only: Used to activate the second measuring range for mass flow.
• Qv only: Used to activate the second measuring range for volumetric flow.
Qm Max DN
Maximum mass flow rate for the selected nominal diameter at a velocity of 10 m/s.
The value is set automatically based on the selected nominal diameter.
Qm Max 1
Used to set the full-scale value 1 (measuring range = 0 to
Qm Max 1
) for the upstream and downstream mass flow.
Default setting: 1 x Q
max
DN
Qm Max 2
Used to set the full-scale value 2 (measuring range = 0 to
Qm Max 2
) for the upstream and downstream mass flow. This
value is also used for determining the corresponding percentage value. This parameter is only available if for the “Qm Range
Mode” parameter setting the “Qm Max 2” value was selected.
Qm Range Mode
Used to manually toggle between the measuring ranges (Qm Max 1 / Qm Max 2) for measuring the mass flow. This
parameter is only available if for the “Range Mode Config” parameter setting the Qm and Qv value or the Qm only value was
selected.
Qv Max DN
Maximum volumetric flow rate for the selected nominal diameter at a velocity of 10 m/s.
The value is set automatically based on the selected nominal diameter.
Qv Max 1
Used to set the full-scale value 1 (measuring range = 0 to
Qv Max 1
) for the upstream and downstream volumetric flow rate.
Default setting: 1 x Q
max
DN
Qv Max 2
Used to set the full-scale value 2 (measuring range = 0 to
Qv Max 2
) for the upstream and downstream volumetric flow rate.
This value is also used for determining the corresponding percentage value. This parameter is only available if for the “Qv
Range Mode” parameter setting the “Qv Max 2” value was selected.
Default setting: 1 x Q
max
DN
Qv Range Mode
Used to manually toggle between the measuring ranges (Qv Max 1 / Qv Max 2) for measuring the volumetric flow. This
parameter is only available if for the “Range Mode Config” parameter setting the Qm and Qv value or the Qv only value was
selected.
Sensor Location Tag
Used to enter the meter location for the sensor.
Alphanumeric, max. 20 characters
Sensor Tag
Used to enter the TAG number of the sensor.
Alphanumeric, max. 20 characters.
...Operating Mode
Used to select the “...Operating Mode” submenu with
.
Содержание 406060
Страница 2: ......
Страница 98: ...13 Appendix 98 13 3 Declarations of conformity 13 3 1 JUMO flowTRANS MAG S10...
Страница 99: ...13 Appendix 99...
Страница 100: ...13 Appendix 100...
Страница 101: ...13 Appendix 101...
Страница 102: ...13 Appendix 102...
Страница 103: ...13 Appendix 103...
Страница 104: ...13 Appendix 104 13 3 2 JUMO flowTRANS MAG H10...
Страница 105: ...13 Appendix 105...
Страница 106: ...13 Appendix 106...
Страница 107: ...13 Appendix 107...
Страница 108: ...13 Appendix 108...
Страница 109: ...13 Appendix 109...
Страница 110: ...13 Appendix 110 13 3 3 JUMO flowTRANS MAG 10...
Страница 111: ...13 Appendix 111...
Страница 112: ...13 Appendix 112...
Страница 113: ...13 Appendix 113...
Страница 115: ......